Remote Control Commands
R&S
®
NRP18S-xx
62
User Manual 1178.3686.02 ─ 05
Parameters:
<level>
Range:
1.0e-7 to 200.0e-3
*RST:
1.0e-6
Default unit: Watts
TRIGger:LEVel:UNIT
<unit>
Sets the unit of the trigger level if this value is entered without a unit.
See also
Parameters:
<unit>
DBM | W | DBUV
*RST:
W
TRIGger:MASTer:PORT
<master_port>
Selects the port where the sensor outputs its own trigger event in case it is trigger mas-
ter. See
for more information.
If the sensor is the trigger master, it can output its trigger event either on the
EXTernal<1>
or
EXTernal2
.
If the sensor triggers itself, the trigger source of the sensor must be assigned to the
other external port, as shown in the examples.
Parameters:
<master_port>
EXT1 | EXTernal1 | EXT2 | EXTernal2
*RST:
EXT1
Example:
TRIG:MAST:PORT EXT1
TRIG:SOUR EXT2
TRIG:MAST:STAT ON
Example:
TRIG:MAST:PORT EXT2
TRIG:SOUR EXT1
TRIG:MAST:STAT ON
TRIGger:MASTer:STATe
<state>
Enables or disables the trigger master mode of the sensor. In this state, the power sen-
sor can output a digital trigger signal in sync with its own trigger event.
If enabled, select the output port for the trigger signal using
.
Typically, the trigger master uses its internal trigger source. But you can also trigger the
trigger master externally, because the power sensor has got two external trigger con-
nectors. If you trigger the trigger master externally, use
EXTernal1
as external trigger
input port (trigger source) and
EXTernal2
as trigger master output port or vice versa.
Parameters:
<state>
*RST:
OFF
Controlling the Measurement