GUI reference
R&S
®
ZNA
457
User Manual 1178.6462.02 ─ 20
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
MINimum | MAXimum
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
Center = Marker
See
"Center = Marker / Start = Marker / Stop = Marker / Span = Marker"
Next Peak
Sets the active marker to the next local maximum or minimum in the search range,
depending on the selected
.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
NPEak
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
Peak Left / Peak Right
Sets the active marker to the next local maximum or minimum to the left or right of the
current marker position, depending on the selected
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
LPEak | RPEak
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
Peak Type
Defines the peak type to be searched for using
:
●
If "Max" is active, then the marker is set to the next maximum. The next maximum
is the maximum with the largest response value that is below the current marker
response value.
●
If "Min" is active, then the marker is set to the next minimum. The next minimum is
the minimum with the smallest response value that is above the current marker
response value.
●
If "Min or Max" is active, then the marker is set to the next minimum or maximum,
whichever has the smallest distance from the current marker response value.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
LPEak | RPEak
Search Range...
Tracking
Enables/disables tracking of the active marker for the current
, which cau-
ses the marker to be updated after each sweep (or after each sweep point in case of
"Sweep Progress").
Among the available search modes, the tracking functionality only makes sense for:
●
"Min" and " Max", where such an update typically causes the marker to change
both its horizontal and vertical position and
●
"Target Search", where typically only the horizontal position changes
●
"Sweep Progress" for long duration sweeps
Marker softtool