Command reference
R&S
®
ZNA
1278
User Manual 1178.6462.02 ─ 20
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DELete:ALL
Deletes all port assignments (see
Chapter 4.5.5.6, "Multiple port assignments"
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:
Usage:
Event
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire
Starts the automatic calibration sweep for the indicated channel and port assignment.
A complete, valid set of port assignments must be defined before you can initiate a cal-
ibration. See
Chapter 4.5.5.6, "Multiple port assignments"
information.
Suffix:
<Ch>
.
Number of the calibrated channel
<Asg>
Number of the port assignment
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:
Usage:
Event
Manual operation:
See
"Start Cal Sweep / Abort Sweep"
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine
<TestPort1>, <CalUnitPort1>[, <TestPort2>]...
Manually defines port assignment no. <Asg> for channel no. <Ch>.
With manual configuration a non-minimal set of port assignments can be created,
which may increase measurement accuracy. On the other hand, it is up to the user to
ensure that the created set of port assignments is complete and valid for the chosen
calibration type (see
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
Note
that during the corresponding calibration sweep the R&S
ZNA expects the physi-
cal port connections to be established exactly as defined by the port pairs.
Use
●
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:
TPORt
command to take advantage of auto-detection of port connections.
●
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:
to create the default assignments instead.