Concepts and features
R&S
®
ZNA
189
User Manual 1178.6462.02 ─ 20
For a channel with more calibrated ports than cal unit ports available, you can change
the port assignments and rerun the automated validation. Or you can switch to
advanced mode and validate each port manually.
Advanced mode
In advanced mode you can configure the validation settings (reference characteriza-
tion, deviation limits) and run or rerun particular measurements (reflection at port p with
connected Open, Short or Match).
4.5.5.6
Multiple port assignments
With multiple port assignments, it is possible to calibrate several test ports that
exceeds the number of ports of the calibration unit or cal unit characterization in use.
E.g. you can use a 2-port calibration unit to perform a full 4-port calibration.
Calibration