Command reference
R&S
®
ZNA
1133
User Manual 1178.6462.02 ─ 20
Parameters:
<Voltage>
Range:
-6 to +12 V
Increment: 0.005 V
Default unit: V
Manual operation:
See
"GPIO columns (sweep sequencer table)"
CONTrol<Ch>:SEQuence<Nr>:RFFE<Bus>:COMMand:DATA
<Command>
For unsegmented sweeps this command allows to define the RFFE command(s) to be
executed at the start of each sweep.
Command execution on an RFFE interface can be enabled/disabled using
CONTrol<Ch>:RFFE<Bus>:SETTings[:STATe]
. Complementary GPIO switches
can be defined using
CONTrol<Ch>:SEQuence<Nr>:GPIO<Port>:VOLTage
Suffix:
<Ch>
.
Channel number
<Nr>
Sequence number, defining the order in which the commands
shall be executed.
For every channel and RFFE interface, the sequence numbers
must be consecutive, starting at 1; reusing a sequence number
overwrites a previously defined command.
Delays between subsequent commands and - finally - the start
of the respective sweep can be introduced using
CONTrol<Ch>:SEQuence<Nr>:DELay
<Bus>
RFFE bus interface number
Parameters:
<Command>
3 to 37
hexadecimal digits
(0-F), defining the command to be
executed: digit 1 is the slave address, digits 2 and 3 specify the
command number and the remaining digits represent the data
part (up to 17 digit
pairs
).
Manual operation:
See
"RFFE columns (sweep sequencer table)"
CONTrol:VSE:ADDRess
<Adress>
Sets/gets the VISA address of a remote R&S
VSE instance.
Parameters:
<Adress>
VISA resource string
Manual operation:
See
CONTrol:VSE:CONFig:LOAD
<FilePathOnVna>
If a connection to a R&S
VSE instance is established, you can load its configuration
from a file on the R&S
SCPI command reference