Command reference
R&S
®
ZNA
1121
User Manual 1178.6462.02 ─ 20
Example:
CONTrol:GPIO:SENSe:SUMCurrent? 1,2,3,4,5
returns the sum of the currents measured in GPIO ports 1 to 5
Usage:
Query only
Options:
R&S ZN-B15/Z15 Var. 03
Manual operation:
See
CONTrol<Ch>:GPIO:SENSe:TRIGger
Starts the voltage/current measurements on all GPIO pins.
The measurement time can be defined using
.
Suffix:
<Ch>
.
Channel number
This suffix is ignored and can be omitted.
Usage:
Event
Options:
R&S ZN-B15/Z15 Var. 03
Manual operation:
See
CONTrol<Ch>:GPIO:TIME
<MeasTime>
Sets the measurement time for the voltage/current measurements on the RFFE and
GPIO pins.
Same functionality as
Suffix:
<Ch>
.
Channel number
This suffix is ignored and can be omitted: the measurement time
is valid for all channels.
Parameters:
<MeasTime>
Measurement (= sampling = averaging) time
Range:
95 μs to 100 ms
*RST:
100 ms
Default unit: s
Options:
R&S ZN-B15/Z15 Var. 03
Manual operation:
See
CONTrol<Ch>:GPIO:VOLTage:OUTPut
Applies the current output voltage and current range (shunt) settings to the RFFE/
GPIO pins.
Suffix:
<Ch>
.
Channel number
SCPI command reference