Command reference
R&S
®
ZNB/ZNBT
884
User Manual 1173.9163.02 ─ 62
Parameters:
<SidebandSupp>
Sideband suppression
Range:
10 dB to 120 dB
Increment: 10 dB
*RST:
32 dB
Default unit: dB
Example:
*RST; :CALC:FILT:TIME:WIND DCH
Reset the instrument and select a Dolph-Chebyshev time gate
for filtering the data in the frequency domain.
CALC:FILT:TIME:DCH 25
Set the sideband suppression to 25 dB.
Manual operation:
See
CALCulate<Chn>:FILTer[:GATE]:TIME:SHAPe
<TimeGate>
Selects the time gate to be applied to the time domain transform.
Tip:
Use the generalized command
CALCulate<Chn>:FILTer[:GATE]:TIME:WINDow
if you wish to select a Dolph-Chebychev time gate.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<TimeGate>
MAXimum | WIDE | NORMal | MINimum
MINimum - Steepest edges (rectangle)
WIDE - Normal gate (Hann)
NORM - Steep edges (Hamming)
Maximum - Maximum flatness (Bohman)
*RST:
WIDE
Example:
*RST; :CALC:FILT:TIME:SHAP?
Reset the instrument and query the type of time gate used. The
response is WIDE.
Manual operation:
See
CALCulate<Chn>:FILTer[:GATE]:TIME:SHOW
<Boolean>
Enables or disables permanent display of the gate limits.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON - time gate permanently displayed
OFF - time gate hidden
*RST:
OFF
SCPI command reference