Physiolaser
olympic 500
Version 1.4 June 2007
38
Laser-protection goggles
Both personnel and patients are to wear laser protection safety goggles during treatment
sessions.
The only types of safety goggles which may be worn are those which meet DIN 58215
standards. The filter glasses effective wavelength range include the laser light
wavelength of the
Physiolaser
olympic
equipment in use. There are two types of
safety goggles available:
Number 530, which may be used for wavelengths from: 655 nm to 904 nm
Cleaning / Disinfecting
Please observe that cleaners for plastics, soap are used for periodic cleaning of the
plastic parts. No penetrative or corrosive agents (such as acetone) should be used.
Keep fluids from entering the housing. Standard disinfectants may be used. When this is
done, the unit is to be wiped with a soft cloth which has been dampened with a mild,
approved disinfectant (50% diluted propyl alcohol solution). The tip may be dipped into
the solution.
Under no circumstances
is the unit to be flooded with the disinfectant or
to be treated with solvents!
Maintenance/Calibration
The laser diodes will be calibrated by the software at each start. RJ recommends an
annual service of the laser at the RJ service facilities or an authorized service station.
Always ensure that operational and sufficiently charged batteries are used.
The interior of the
Physiolaser
olympic
contains no elements or components which
require servicing by the user.
Therefore, there is no need to open the housing. To meet the requirements of DIN VDE
0837, “Radiation safety for laser devices,” the unit is to be tested and recalibrate after
one year of use to ensure that the laser is operating properly. Please send the unit to the
manufacturer for testing and recalibration.
Disposal
Due to the hazards of laser beams, the
Physiolaser
olympic
is not to be disposed of
as standard electronic scrap. The device is to be delivered to either the supplier or to the
manufacturer for proper disposal.
Warranty
There is a 2 years warranty from the manufacturer. The manufacturer guarantees
operation of all
Physiolaser
olympic
features described in the user’s guide, provided
the equipment is used in accordance with the instructions detailed in the user’s guide
and is treated with proper care.
New settings, maintenance or repairs may only be performed by the manufacturer or an
authorized workshop; otherwise the warranty is rendered null and void. Interventions in
the device by third parties will lead to loss of warranty claims and will release the
manufacturer from all product liability.
The manufacturer will keep spare parts in stock for at least 10 years after the purchase
date.