36
before
the
deflation
step.
Test
the
cuff
on
your
finger
to
observe
the
deflation
steps
to
determine
whether
the
cuff
or
hose
is
leaking.
Try
another
cuff
if
in
doubt
to
confirm
proper
operation.
Discard
any
leaking
cuff,
as
using
a
leaking
cuff
will
result
in
inaccurate
readings.
Cuff
will
not
inflate.
1.
Check
cuff
connections,
both
at
the
cuff
port
of
the
device
and
any
connections
between
the
cuff
and
the
extension
hose.
Check
cuff
and
extension
hose
for
leaks
or
damage.
2.
Verify
that
the
pump
is
actuating
(you
will
hear
the
noise
and
feel
a
slight
vibration)
when
BP
Start/Stop
is
pressed.
If
not,
contact
Customer
Service.
Screen
buttons
don’t
respond
in
the
expected
way
to
touch.
It’s
possible
that
the
touchscreen
needs
recalibration.
To
do
so:
1.
Turn
the
unit
OFF.
2.
To
enter
the
setup
mode,
press
your
thumb
or
finger
gently
on
the
screen
and
then
turn
the
unit
ON
while
continuing
said
pressure
on
the
screen.
When
the
counter
in
the
upper
left
reaches
exactly
8,
release
your
thumb
or
finger
from
the
screen
and
follow
the
screen
instruction
(See
Figure
1).
3.
Follow
the
prompts
on
the
screen
by
touching
the
arrows
with
a
pointed
object
like
a
stylus
or
pencil.
Do
not
use
a
sharp
pointed
object
in
order
to
avoid
damage
to
the
display
(See
Figure
2).
4.
The
unit
will
acknowledge
each
successful
touch
by
beeping
and
counting
down
before
moving
to
a
different
screen
location.
5.
Once
all
three
locations
have
each
been
touched
three
times,
touch
the
screen
anywhere
to
exit
the
screen
calibration
function.
6.
When
the
unit
exits
the
setup
mode
and
returns
to
its
normal
operating
mode,
verify
that
the
touch
buttons
work
as
expected.
Figure
1
Figure
2