Default settings are for a LOW alarm tone pitch, device supervision turned OFF, and direct signaling ON. This will allow the pager to
be paired 1-to-1 with a Wireless nurse call handset or with a Wireless bed, seat or floor sensor pad
To enter/exit programming mode, press and hold the Option button for ~3s. While the pager is in programming mode, the display
will show a two-letter prefix on the left side to indicate the setting that is to be adjusted, with the current setting shown on the right
side of the screen.
1.
Read-in Devices (rd)
This is used when you want to add or delete a Wireless pad, wristband transmitter, or other wireless device from the pager memory.
The pager has memory for maximum 9 devices. If you have more than 9 devices in an area you should instead use the optional
central monitor.
In the programming menu select “rd”
•
Press Option to enter the menu, and a digit from 1-9 will start flashing in the center of the screen
•
Use Menu to choose a memory slot from 1-9
•
The right side of the LCD will indicate the memory slot is Open (O) or Full (F)
•
Trigger wireless device to add into the memory slot – the LCD will update to show ‘F’
•
Hold Cancel ~3 to remove device from memory slot – the LCD will update to show ‘O’
•
Press Option again to exit pairing and return to programming menu
Use the Menu button to navigate the pager options, looking at the LCD to choose the appropriate setting from the above table to
adjust. Pressing the Option button will allow the current setting to be adjusted, and it will begin flashing on the LCD. Use Menu to
adjust the setting, and confirm by pressing Option again.
3. Programming
Wireless Jill – Pager Setting
LCD Indicator
Read-in Devices
rd [1-9] [O/F]
Select Tone
to [LO/HI]
Direct Signaling
dS [On/OFF]
Supervision
Su [On/OFF]
Reset
rE [0-1]
Site Code
SC [0-7]
Clear Memory
Cl