Validation of Specification Data
• Validation of minimum filtration of the radiation unit (x-ray tube, collimator and possible fil-
ters) must be performed on a completely installed system by whoever is responsible for
the installation.
• Validation of light field luminosity and contrast is performed in-factory. See Chapter -
C
OM
-
PLIANCE
V
ERIFICATION
,
PARAGRAPH
:
L
IGHT
F
IELD
I
LLUMINATION
I
NTENSITY
.
• Validation of system x-ray leakage is to be performed following the installation of the sys-
tem components under the responsibility of the person performing the installation.
• The correct installation of the system, electronic calibration and wiring layout of the same
are the responsibility of the engineer responsible for the installation of the system.
• Protection of the cable is to be provided by the installation engineer.
Ralco is prepared to provide any information required regarding the validation methods
described above.
EN 60825-1 par.1 -5
The collimator laser is classified as Class 2M; the light-field led is classified as Class 2 and is
used for light field projection.
Alimentazione scheda esterna
n.a.
Power supply, external PCB
Fusibile per scheda esterna
n.a.
Fuse for external board
CLASS 2 LED PRODUCT=> WHITE LED
CLASS 2M LASER PRODUCT=>RED LASER LINE