MTR302/A - R302F/A - R302L/A DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
Modello/Model R302/A
Substitutions
1
R302/A
- Substitutions
Ralco
srl
SUBSTITUTIONS
Substitution of the Led
• Disconnect supply.
• Remove the cover - See Chapter -
C
OVER
R
EMOVAL
.
• Remove the Led protection heatsink.
• Unsolder two wires.
• Unscrew two fixing screws.
• Remove the faulty Led.
• Substitute the Led with an identical item; handle it gently and make sure it doesn’t get dirty.
• Tighten the two fixing screws.
• Solder the two wires.
• Check on light field/x-ray field correspondence.
Fig. Led substitution
Substitution of the Electronic Board for Power Led
• Disconnect Supply.
• Remove part of the cover to access the component - See Chapter
- C
OVER
R
EMOVAL
.
• Remove the two screws holding the electronic timer.
• Identify the cables and their position on the terminal board. See Chapter -
E
LECTRICAL
C
ONNECTION
.
• Disconnect the cables from the terminal board.
• Install the new timer by proceeding in a reverse order; pay particular attention to the con-
nection of the cables on the 2 or 8-way electric terminal.
Substitution of the Laser
• Disconnect supply.
• Remove the cover - See Chapter -
C
OVER
R
EMOVAL
.
• Unscrew the fixing Allen screws on the upper part of the support - See Chapter -
A
DJUST
-
MENTS
.
• Disconnect the timer cables from the terminal board - white 0V, black 5V.
• Remove the laser and substitute with an identical item.
• Tighten the screws.
• Check on laser alignment - See Chapter -
A
DJUSTMENTS
.
• Remount the cover - See Chapter -
C
OVER
R
EMOVAL
.