Modello/Model R302/A
Calibration
3
R30
2/A
- Cali
bration
Ralco
srl
• Use screw
B
to adjust transversally.
• When calibration is terminated, lock the screws
A.
See the
Figure
that follows.
V
ERTICAL
A
LIGNMENT
• Remove the part of the cover necessary to access the screws - See Chapter -
C
OVER
R
EMOVAL
.
• If adjustment is required loosen the two screws
C
holding the led support.
• If the light-field is smaller than the x-ray field, lower the led by adjusting screw
D
.
• If the light-field is bigger than the x-ray field, raise the led by adjusting screws
D
.
• Tighten the two screws
C
.
See the following
Figure
.
Fig. Light field - x-ray field adjustment
Transversal Calibration (CROSS)
If the light-field needs calibration, the mirror needs to be adjusted as follows:
• Remove the part of the cover necessary to access the screws - see Chapter -
C
OVER
R
EMOVAL
.
• Loosen the mirror fixing screw
- A -
and adjust cam -
B
- just enough to be able to rotate the
cam and position the mirror. See the
Figure.
• Once you have regulated the mirror tighten the screw -
A-
and remount the cover - see
Chapter -
C
OVER
R
EMOVAL
.