2
www.raesystems.com
Buzz & Lite
Set the behavior of the audible and visible alarms.
Choices are Both on (LEDs and audible alarm), Light
only (LEDs only), or Both off.
Note:
You cannot turn off all of the alarm alerts at the
same time. If you select Both Off when the Vibration
alarm is Off, “Invalid Setting” is displayed and the
previous Buzz & Lite setting is restored.
Vibrate Mode
Set the vibration alarm. Choices are On or Off.
Note:
You cannot turn off all of the alarm alerts at
the same time. If you select Both Off when the Buzz
& Lite is set to Both Off, “Invalid Setting” will be
displayed and the previous Vibrate setting will be
restored.
Temp Unit
Set the unit of measure. Choices are Fahrenheit and
Celsius. Press MODE to select Fahrenheit or Celsius.
Then press SET to save your choice and exit.