OPERATION
2-16
2.5
Alarm Signals
During each measurement period, the gas concentration is
compared with the programmed alarm limits (gas concentration
alarm limit settings: Low, High, TWA and STEL). If the
concentration exceeds any of the preset limits, the loud buzzer
and red flashing LED are activated immediately to warn the
user of the alarm condition.
In addition, the MiniRAE 2000 will alarm if one of the
following conditions occurs: battery voltage falls below a pre-
set voltage level (4.4 V), failure of UV lamp, pump stall, or
when the datalog memory is full. When the low battery alarm
occurs, there will be approximately 20-30 minutes of operating
time remaining. When the battery voltage falls below 4.2 V, the
monitor will turn off automatically.