OPERATION
2-9
The Main Menu Functions
•
Ready…:
Indicates that the monitor is ready to take a
measurement or to step through the Main Menu. Press the
[Y/+]
button to advance to taking a measurement (read
“Taking a Measurement” on Page 2-12 for details).
Note: The
Ready…
screen is skipped if the menu is cycled
through while a measurement is running.
•
Check…:
This message displays while the system is
checking the lamp and the sensor. If the lamp test succeeds,
the system will progress to checking the sensor. If the lamp
test does not succeed, the display will read
Can not turn on
lamp, quit!
-
Can not turn on lamp, quit!:
This message displays
when the lamp does not turn on. The system will
automatically return to
Ready…
allowing the user to
test the lamp again. If the lamp fails a second time, turn
the monitor off and refer to Section 7.2 “
PID Sensor &
Lamp Cleaning / Replacement
”.
-
Clean Sensor, Quit?:
This message displays when the
sensor requires cleaning. Press the
[Y/+]
button to turn
the monitor off and clean the sensor. Press the
[N/-]
button and the system will progress to the
Site ID = xx
•
Site ID = xxxx:
This display shows the Site ID and
indicates that the monitor is about to start taking
measurements (read “Taking a Measurement” for details)
Note: This display appears only after a measurement has
been started. It does not appear when the user is cyling
through the Main Menu and the monitor is idling.
•
Gas = xxxx:
This display identifies the gas to be measured
and indicates that the monitor is about to take a
measurement (read “Taking a Measurement” for details)