- 48 -
27.4.7. Send Stable Measurement Result in Current Measuring Unit
Format:
SU CR LF
Response options:
SU_A CR LF
SU_E CR LF
- command understood and in progress.
- time limit exceeded while waiting for stable measurement result.
SU_I CR LF
- command understood but not accessible at this moment.
SU_A CR LF
MASS FRAME
- command understood and in progress.
- response: mass value in current measuring unit.
Response format:
1 2
3
4
5
6
7-15 16 17 18 19 20 21
S U space
stability
marker
space character mass space
unit
CR LF
Example:
S U CR LF
– command sent from a computer.
S U _ A CR LF
- command understood and in progress.
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF
- command carried out, response:
mass value in current measuring unit.
27.4.8. Immediately Send Measurement Result in Current Measuring Unit
Format:
SUI CR LF
Response options:
SUI_I CR LF
- command understood but not accessible at this moment.
MASS FRAME
- immediate response: mass value in current measuring unit.
Response format:
1 2 3
4
5
6
7-15 16 17 18 19 20 21
S U I
stability
marker
space character mass space
unit
CR LF
Example:
S U I CR LF
– command sent from a computer.
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF
- command carried out, immediate
response: mass value in current measuring unit.