The tare value acquired from the weighing device memory is
not remembered upon the weighing device restart.
13.8. Last Digit
Function designed to disable display of the last weighing indication digit, this
results with less accurate measurement.
Procedure:
•
Enter
<P2.rEAd / 2.8.LdiG>
submenu.
•
Press
key, available values are displayed successively one by one:
ALAS
Select to make the last digit always on.
nEur
Select to make the last digit always off.
uuSt
Select to make the last digit on only when the weighing indication is stable.
•
Press
key to confirm, next go to the home screen.
14. COMMUNICATION
Communication between the scale and the peripheral devices is established
via the following ports: RS232, USB type A, USB type B. To set the ports go to
<P4.Conn>
submenu.
14.1. RS232 Port Settings
•
Enter
<P4.Conn / 4.1.rS1>
submenu and set respective transmission
parameters:
4.1.1.bAd
Baud rate: 2400, 4800, 9600, 19200, 38400, 57600, 115200 bit/s.
4.1.2.PAr
Parity:
nonE
– none;
EuEn
– even;
Odd
– odd.
•
Press
key to confirm, next go to the home screen.