To enable the function in parameters you need to:
•
Enter parameter group
<
Working modes>
according to ch.
22 of this manual, choose:
„
Counting Pieces /
Automatic
correction of reference mass”
and set appropriate option.
Options:
No
- Automatic correction of reference mass disabled
Yes
- Automatic correction of reference mass enabled
Function
<
Automatic correction of reference mass>
in mode
<
Counting Pieces>
is enabled at the moment of estimating the sample
quantity and signalled by displaying
<PCS>
and
<SMP>
(single piece mass)
on the top part of the display.
There are four criteris of working “Automatic correction of reference
mass”
function:
1. equilibrium should be reached,
2. quantity of pieces should be increased,
3. added quantity of pieces should not be greater than double
number of pieces on the pan,
4. the new sample can be different from the old sample by
±
0.3
of pcs (absolute value),
If a user recognises that the sample quantity is adequate the unit mass
(single piece mass) can be saved (see ch. 24.5 of this manual)
and disable the function by pressing
.
Notice:
While the function is active
changes its functionality.
Pressing
does not result in printing and saving weighings.