Put a weight of the displayed mass value on the pan and press
.
The calibration process will start which is signalled by the message:
•
After completion of the process of calibration the following screen
will appear
•
Take off the weight , then the following sequence of screens
will appear
•
Calibration process can be terminated anytime by pressing
which is signalled by the following message on the display:
•
Return to weighing with saving changes that have been made.
Caution:
If the calibration process (span adjustment) lasts longer than 15 the
<Err8> message will be displayed and short audible signal will be
emitted. Press
to perform calibration again with more stable
ambient conditions!
- 61 -
Содержание WPF
Страница 2: ...2 JANUARY 2012...
Страница 11: ...WPT H1series pillar 7cm main dimensions WPT H2 WPT H6 series main dimensionS 11...
Страница 13: ...WPT C1 R series main dimensions WPT C1 series main dimensions 13...
Страница 14: ...WPT C2 K series main dimensions WPT C2 R series main dimensions 14...
Страница 15: ...WPT C2 series main dimensions WPT C3 K series main dimensions 15...
Страница 35: ...Return to weighing see 10 2 2 13 3 Baud rate Procedure Enter the submenu P2 Prnt and then 35...