14.2. “Beep” signal – after pressing a key
Procedure:
•
Enter the submenu
<P5.othr>
and then:
bEEP
no
- switched off
bEEP
YES
- switched on
Return to weighing:
See 10.2.2.
14.3. Automatic switch-off
This function is essential to save the battery power. The scale is switched
off automatically when (function
t1 = YES
) no weighing appears in 5
minutes. (no changes on the display). In case when this function disrupts
the operation (e.g. long time weighing procedures) or while working with
connection to mains, switch off this function.
Operation according to the power supply:
Operation
Setting
Mains Batteries/accumulator
t1 = 0
disabled
disabled
t1 = YES
enabled
enabled
t1 = Auto *
disabled
enabled