11.3. Setting the time of display refreshment
This parameter determines how often indication of display is refreshed. It is
realized by comparing what in given moment is on the display with information
sent by balance processor about state of weighed mass placed on the pan.
For higher refreshment parameters value display do not indicate intermediate
unstable mass indication by placing and removing the load on balance pan,
however for low values all changes of mass during weighing are visible – e.g.
this enables dosing loose or liquid materials. Time of refreshment is set in
seconds.
11.4. Setting of autozero operating
To ensure accurate balance indications software “AUTOZERO” function was
introduced. Task of this function is automatic control and correction of balance
zero indication.
When function is active comparison of successive results takes place in
declared distance of time e.g. every 1 s. If these result differ in value lower than
declared range of AUTOZERO e.g. 1 interval, balance will automatically zero
and markers of stable result –
and zero indication –
will be
displayed.
When AUTOZERO function is on every measurement starts from precise zero.
There are some special cases when this function interrupts in measurements.
Example of it can be very slow placing the load on the pan (e.g. pouring the
load) in such case system of zero indication correction can also correct
indication of real mass of load. AUTOZERO can be switch on or off in
parameter P4 06 – main platform 07 – additional platform.
11.5. Last digit
To ensure appropriate comfort of operating with balance user determines
(depending on needs) if last digit of mass indication is to be displayed and in
which states of balance. One of below values can be set:
-
0 never
-
1
always
-
2
when stab
For verified balance option 1 – always is set (without possibility of changing)