26
11. SETTING WORK PARAMETERS
User by means of appropriate setting of parameters from menu group <P4
Readout> can adjust the balance to existing operating conditions (filter) and
one’s expectations (refreshing, autozero, last digit displaying)
Fig. 13. Submenu Readout – internal settings
11.1. Filter setting
For perfect conditions filter can be set as very fast, however if conditions are
bad (vibrations, draughts) filter should be set as slow or as very slow.
Effectiveness of filter operating is different for weighing range. Filter works
slower in time of “reaching” the final result, however stronger when mass gets
to set range of filter operating (parameter range of filter operating available only
in service menu – user do not have access to it).
11.2. Median filter setting
Task of median filter is to eliminate single big disturbances. By setting digital
value determines speed of filter operating. For zero option filter operating is off.