39
Use shielded cables of wire cross-section ranging from 0.25mm
2
to 1mm
2
.
While selecting load cell, it is necessary to make sure that
requirements regarding intrinsic safety parameters are met
for measurement inputs/outputs 5V, E, AGND,
REF+, REF-, +IN, -IN.
14.4. 4-Wire Load Cell Connection
Connect 4-wire load cell to the main board following the diagram below:
4-wire load cell connection
466R RADWAG board marking
Load cell signals
SHIELD
IN+
IN- OUTPUT-
+5V INPUT+
AGND INPUT-
Use shielded cables of wire cross-section ranging from 0.25mm
2
– 1mm
2
.
Содержание PUE HX5.EX