9.2. Adjustment test
Adjustment test is a comparison of internal adjustment weight with its value stored in
balance’s memory. This process is carried out automatically and its result is shown on the
display.
Fig. 15. Adjustment test
Cal. – the value of internal adjustment weight
Curr. – weighing result of the internal adjustment weight
Dev. – deviation calculated between the both values
Return to weighing mode
Changes introduced in balance memory will be saved on returning to
weighing with procedure of saving changes. Press ESC key for a few times
until the display shows a question: Save? As displayed, select one of
available options:
ENTER – save changes and go back to menu;
ESC
– abandon changes and go back to menu.
(see point 7.1.2. Return to weighing)
9.3. Manual adjustment
9.3.1. Internal adjustment
Press CAL key on balance’s overlay, or
1. Go to submenu P1 – Adjustment.
2. Place the marker next to a function 01 Internal adjustment.
3. Press
RIGHT ARROW KEY
.
4. The balance automatically carries out internal adjustment process. While
adjustment process is in progress do not load the weighing pan with any weight.
5. On completing the internal adjustment procedure the balance saves adjustment
data in its memory and returns to weighing mode.
Caution:
-
In order to abort adjustment process press ESC key.
-
If during internal adjustment process the weighing pan is loaded, then the
balance displays an error message. The adjustment process is automatically
stopped. On taking off the load from the weighing pan, the process is resumed
and completed.
-
If the DRH function is enabled in balance settings, then balance user cannot
abort the adjustment process once initiated.