- 169 -
21. COMMUNICATION PROTOCOL
General information
A. A character based communication protocol balance-terminal is
designed for establishing communication between a RADWAG
balance and peripheral devices via RS 232 interface.
B. It consists of commands sent from a peripheral device to the balance
and responses from the balance.
C. Responses are sent from the balance on each receipt of a command
as a reaction to a specific command.
D. Commands forming the communication protocol enable both, obtaining
data on balance status and influencing its operation, e.g.: acquiring
measurement results from the balance, zeroing, etc.
21.1.
List of commands
Command Command overview
Z
Zero balance
T
Tare balance
OT
Give tare value
UT
Set tare
S
Send stable measurement result in basic measuring unit
SI
Immediately send measurement result in basic measuring unit
SU
Send stable measuring result in current measuring unit
SUI
Immediately send measurement result in current measuring unit
C1
Switch on continuous transmission in basic measuring unit
C0
Switch off continuous transmission in basic measuring unit
CU1
Switch on continuous transmission in current measuring unit
CU0
Switch off continuous transmission in current measuring unit
DH
Set low checkweighing threshold
UH
Set high checkweighing threshold
ODH
Give value of low checkweighing threshold
OUH
Give value of high checkweighing threshold
SM
Set mass value of a single item