In order to accept carried out changes, press
Print/Enter
key for three (3)
times.
Parameter
active drying mode
can take the following values:
•
Drying mode
STANDARD
In standard mode, user should set drying temperature
tmP
in which a sample
should be tested,
•
Drying mode
QUICK
In quick mode, user should set drying temperature
tmP
in which a sample
should be tested,
Specific feature of this drying mode is fast increase of drying temperature in
short period of time. Drying temperature for the time of 180 seconds is
exceeding the pre-set value by 30 %, after which is drops to temperature value
set in the parameters of the drying mode.
•
Drying mode
MILD
In mild mode, user should set drying temperature
tmP
in which a sample
should be tested, and time
t1
in which a moisture analyzer should reach pre-
set temperature value.