background image

 

 

 

User manual no.: 
LMI-51-05/03/13/A 

 

 

 
 

 

 

 

 

 

 

 

 

 

MOISTURE  
ANALYZER  
MAC 
SERIES 

 

 

 

 

 

 
 
 
 

 

 
 
 

 

 

B A L A N C E S   A N D   S C A L E S  

RADWAG 26 – 600 Radom, Bracka 28, POLAND 

Phone: +48 (0-48) 38 48 800, fax. +48 (0-48) 385 00 10 

[email protected] 
www.radwag.com 

Содержание MAC Series

Страница 1: ...anual no LMI 51 05 03 13 A MOISTURE ANALYZER MAC SERIES B A L A N C E S A N D S C A L E S RADWAG 26 600 Radom Bracka 28 POLAND Phone 48 0 48 38 48 800 fax 48 0 48 385 00 10 export radwag com www radwa...

Страница 2: ...2 MARCH 2013...

Страница 3: ...alyzer temperature stabilization period 14 7 USER MENU 15 7 1 Moving through user menu 15 7 2 Return to weighing mode 17 8 WEIGHING 18 9 MOISTURE ANALYZER MAIN PARAMETERS ADJUSTING MOISTURE ANALYZER O...

Страница 4: ...sent during drying process 37 14 DRYING PROCESS WITH USE OF SHORTENED MENU 38 14 1 Setting means of finish mode 39 14 2 Active drying mode 40 14 3 Type of displayed measurement result 43 15 DRYING PRO...

Страница 5: ...moisture content reading 0 24 sample mass up to 2g 0 06 sample mass up to 2 10g 0 04 sample mass over 10g Drying temperature range max 160 C model NH NP and NS max 250 C model WH Heating module halog...

Страница 6: ...in accordance with safety principles of low voltage instruments A symbol visible on instrument s top cover see below is a warning against its hot surface 2 1 Intended use A moisture analyzer MAC serie...

Страница 7: ...zer that get hot while drying procedure i e disposable pan pan handle and inner shields of the drying chamber Remember that some of tested samples may become dangerous if heated appearance of poisonin...

Страница 8: ...out in low temperatures remember that aggressive substances may cause corrosion to the device 2 4 Warranty Warranty does not cover the following cases Not observing the regulation listed in user manu...

Страница 9: ...yzer Before packing disconnect all attached cables and remove any loose movable parts weighing pan shields inserts Place moisture analyzer and its components in their original packaging and protect th...

Страница 10: ...cts as part of the balance is a strong magnet 4 2 Unpacking Carefully remove the moisture analyzer from its packaging remove the plastic and foil transport protective elements Gently place the moistur...

Страница 11: ...rel use most gentle movements not to damage moisture analyzer s measuring system 4 3 Level setting Before plugging to mains level the moisture analyzer using adjustable feet Turn the feet in a way tha...

Страница 12: ...r s power plug is located at the back of instrument s housing On plugging to mains moisture analyzer s displays software s name and number after which the indication changes to 0 000 g in case of a mo...

Страница 13: ...ge of parameters value moving in moisture analyzer s menu Print Enter key sending display status to peripheral device Print or confirmation of selected value or function Enter ZERO TARE key zeroing mo...

Страница 14: ...lly stabilized It is a period of so called moisture analyzer s self heating For moisture analyzer MAC series which before plugging to mains was stored in room temperature self heating period takes app...

Страница 15: ...wards changing parameter value by one value upwards Navigating key DOWN Selecting group of parameters one by one downwards changing parameter value by one value downwards Navigating key RIGHT Selectin...

Страница 16: ...tout of user name P2 06 PPrJ YES no printout of project name P2 07 PId YES no printout of moisture analyzer s factory no P2 08 PFr YES no printout of frames P3 tinnE Setting time and date of the moist...

Страница 17: ...s P6 05 PrnS printout of moisture analyzer parameters 7 2 Return to weighing mode CAUTION Changes introduced in moisture analyzer memory will be permanently saved on return weighing with procedure of...

Страница 18: ...ult Mass indication of a load placed on moisture analyzer s weighing pan can be tarred for multiple times by pressing ESC TARE key pay attention not to exceed maximal capacity of a moisture analyzer b...

Страница 19: ...according to user needs autozero operation or display backlight by using parameters grouped in this function 9 1 Setting filtering level use navigating keys to select desired filter value AuE StAnd no...

Страница 20: ...d and display indication is close to zero If results vary less than declared AUTOZERO range e g one division the moisture analyzer will zero automatically display marker of stable measurement result a...

Страница 21: ...9 3 Access to library of drying programs In order to use the library of drying programs set parameter Libr to value YES no libraries of drying programs disabled YES libraries of drying programs enabl...

Страница 22: ...bled automatically if measured indication does not change within 10 seconds CAUTION Moisture analyzer software has implemented option of automatic backlight switch off bl Aut In case of this setting t...

Страница 23: ...23 9 5 Brightness adjustment of display backlight 100 maximum display brightness 20 minimum display brightness no display brightness disabled...

Страница 24: ...24 9 6 beep sound reaction on pressing a function key bEEP On signal of pressing function key enabled bEEP OFF signal of pressing function key disabled...

Страница 25: ...eighing accuracy it is recommended to periodically refresh moisture analyzer s accuracy by carrying out adjustment process Adjustment should be carried out Before the beginning of measuring process If...

Страница 26: ...icates mass of a standard weight to be loaded on the weighing pan on loading the mass standard weight press Print Enter key On completing adjustment process the moisture analyzer returns to submenu P1...

Страница 27: ...his process is automatic and its result is previewed on the display if a moisture analyzer is connected to a computer or a printer through RS 232 interface data on adjustment test is printout only if...

Страница 28: ...amber adjustment is a process aimed at adjusting the temperature sensor of a moisture analyzer To carry out temperature adjustment of a moisture analyzer place a control thermometer in an opening of t...

Страница 29: ...to be equal to the current temperature displayed on the control thermometer Use Print Enter key to start part two of temperature adjustment process The moisture analyzer switches on the halogen lamp a...

Страница 30: ...as in previous steps On entering temperature value accept it by pressing Print Enter key Temperature adjustment process is completed and the display indicates name of the parameter P1 03 tE_CAL In cas...

Страница 31: ...MINING CONTENT OF A PRINTOUT FOR GLP PROCEDURES Group of parameters P2 GLP enables declaring variables that are present on an adjustment printout and printout activated after a measurement P2 GLP P2 1...

Страница 32: ...alents displayed by the moisture analyzer An instance of a user name inserted to a moisture analyzer using capital letters takes below form WILK An instance of a user name inserted to a moisture analy...

Страница 33: ...analyzer using small letters takes below form polon P2 03 Ptin Option determining presence of measurement time on a printout P2 04 PdAt Option determining presence of measurement date on a printout P2...

Страница 34: ...on a printout YES variable present on a printout Return to weighing mode see point 7 2 return to weighing mode 12 DATE AND TIME SETTINGS 12 1 Setting time This parameter enables setting time in a moi...

Страница 35: ...nables setting date in a moisture analyzer In order to change date settings use navigating keys Up Down Left Right 13 RS 232 FUNCTIONS 13 1 Baud rate setting Select one of available baud rates 2400 bi...

Страница 36: ...7 2 return to weighing mode 13 2 Determining data type sent via RS 232 output YES sending stable measurement result no sending stable or temporary unstable measurement result Return to weighing mode s...

Страница 37: ...vailable settings 1 2 3 5 10 20 30 60 120 180 seconds 13 4 Determining range of data sent during drying process YES data limited to drying process result only printout of a single like no printout of...

Страница 38: ...ble access to libraries of drying programs go to menu P6 01 Libr In order to change parameter settings enter the drying menu by pressing Test menu key Marker of submenu group drying process finish mod...

Страница 39: ...mass change 1mg in 10s time 2 automatic finish mass change 1mg in 25s time 3 automatic finish mass change 1mg in 60s time 4 automatic finish mass change 1mg in 90s time 5 automatic finish mass change...

Страница 40: ...40 In order to accept carried out changes press Print Enter key for three 3 times 14 2 Active drying mode...

Страница 41: ...ng temperature tmP in which a sample should be tested Specific feature of this drying mode is fast increase of drying temperature in short period of time Drying temperature for the time of 180 seconds...

Страница 42: ...e STEP In step mode user should set drying temperature tmP in which a sample should be tested temperature values tmP1 and tmP2 and time values t1 and t2 between reaching the temperature values in each...

Страница 43: ...ype of displayed measurement result In order to accept carried out changes press Print Enter key for two times Percent loss of mass Displays mass change recorded during drying process and expressed in...

Страница 44: ...lt is mass change recorded by the moisture analyzer in the drying process 15 DRYING PROCESS WITH USE OF DRYING PROGRAM LIBRARY In order to start drying process with use of drying program library the u...

Страница 45: ...l When enabling the option a moisture analyzer software informs the user whether the mass to be tested is within set thresholds regarding the mass and tolerance values on condition that the parameters...

Страница 46: ...mass control Sample mass control option is active only in case of enabled access to library of drying programs P6 01 Libr parameter set to YES The mass control value and tolerance value are set in the...

Страница 47: ...ect a drying mode and then act according to below guidelines Start drying process by pressing Start Stop key and on emptying moisture analyzer s drying pan pressing TARE key Open the drying chamber s...

Страница 48: ...48 Changing of displayed data during drying process is carried out by pressing Display key...

Страница 49: ...of drying process by pressing Print Enter key An exception from the above procedure is means of aborting the drying process in manual mode on pressing Start Stop key the moisture analyzer aborts dryi...

Страница 50: ...NDS FROM THE INSTRUMENT Command PC CR LF sends information on all implemented commands in the program from the instrument Function SEND STABLE RESULT IN BASIC UNIT Command S CR LF sends result in basi...

Страница 51: ...N IN BASIC UNIT Command C1 CR LF starts continuous transmission in basic unit Function GIVE SERIAL NUMBER OF THE INSTRUMENT Command NB CR LF sends factory number Function GIVE WEIGHING RANGE Command F...

Страница 52: ...bility status in set time Er7 ou Inserted value is out of range reg setting reference mass date etc Error 1 3 AD Th L Error 1 3 AD Th H The value of A D divisions of the thermometer in the weighing ch...

Страница 53: ...NUFACTURER O F E L E C T R O N I C W E I G H I N G I N S T R U M E N T S RADWAG BALANCES AND SCALES POLAND 26 600 Radom Bracka 28 Phone 48 48 38 48 800 fax 48 48 385 00 10 export radwag com www radwag...

Отзывы: