- 73 -
The frame format sent by the indicator in case of setting
<P2.Prnt>
to
Cntb
:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
U
I
stability
space
sign
mass
space
unit
CR
LF
20.7. Configuring printouts
General information
If some information included are redundant or not sufficient and there
is a necessity of changes one can design their own protocol format in
EDYTOR WAG
computer program.
This piece of software is accessible
in: http://www.radwag.com
21. ERROR COMMANDS
Err2
- Value beyond the zero range
Err3
- Value beyond the tare range
Err4
- Calibration mass or start mass beyond the acceptable
range (
1% for weight,
10 for start mass)
Err5
- Mass of a single piece lower than the scale division
Err8
- Exceeded the time for tarring, zeroing, start mass
adjustment or span adjustment
NULL
- Zero value from the AD converter
FULL2
- Measurement range
overflow
LH
- Start mass error, the mass on the weighing platform is
beyond the acceptable range (-5% to +15% of start mass)
5-FULL
- Display range overflow in totalizing
Notice:
1. Errors:
Err2
,
Err3
,
Err4
,
Err5
,
Err8
,
null
, that appear on the display
are also signalled by a short beep sound (about 1 sec.);