44
Format:
SUI CR LF
Response options:
SUI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- immediate response: mass value in basic measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
U
I
stability
marker
space
stability
mass
space
unit
CR
LF
Example:
S U I CR LF
- command sent from a computer
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF
- command carried out, immediate
response: mass value in basic measuring
unit
where:
_
- space
Switch on continuous transmission in basic measuring unit
Format:
C1 CR LF
Response options:
C1_I CR LF
- command understood but not accessible at this moment
C1_A CR LF
- command understood and in progress
MASS FRAME
- command carried out, immediate response: mass value in basic
measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
I
space
stability
marker
space
character
mass
space
unit
CR
LF
Switch off continuous transmission in basic measuring unit
Format:
C0 CR LF
Response options:
Содержание HRP 16
Страница 1: ...HRP Weighing Platforms SERIES High Resolution Platforms IMMU 31 07 07 18 EN USER MANUAL www radwag com...
Страница 2: ...2 JULY 2018...
Страница 4: ...4 1 GENERAL INFORMATION 1 1 Dimensions HRP 16 32 HRP 62 120...
Страница 5: ...5 HRP 150 300 HRP 300 1 600 1100...
Страница 6: ...6 HRP 2000 1 2 Connectors Arrangement HRP 16 32...
Страница 7: ...7 HRP 62 120 HRP 150 300 300 1 600 1100 HRP 2000...
Страница 18: ...18 3 Install weighing pan receivers fit them into lever seats...
Страница 26: ...26 4 Disassemble transport lock by undoing screws and removing fixing washers 6...
Страница 28: ...28 10 Connect the scale to the power supply...
Страница 33: ...33 11 Connect the scale to the power supply...
Страница 53: ...53 RS232 HRP HY10 PUE 5 cable power supply Ethernet HRP HY10 PUE 5 cables...
Страница 54: ...54 Ethernet HRP Ethernet Switch Pue 7 1 cable HRP IN OUT cable...
Страница 57: ...57...
Страница 58: ...58...