8. BASIC OPERATIONS
Wait for the adjustment completion.
1
2
Press key to start the adjustment.
8.1.1. Adjustment (for models with an internal adjustment only)
8.1.2. Adjustment (for models with an external adjustment only)
1
2
Press to enter “Setup” menu.
To proceed to calibration, press key.
The adjustment mass standard does not come with
the balance. Prior to adjustment prepare the right
mass standard.
You are recommended to use mass standard
of class F1 or F2. View the table and select
the mass standard you need.
It is recommended to carry out adjustment operation prior to running the WEIGHING mode. External adjustment
is carried out using an external mass standard of the right accuracy and weight value, which value depends
on device type and capacity.
Balance
model
Mass
standard
[g]
PS 200/2000.R1
1 000
PS 600.R1
500
PS 750.R1
500
PS 1000.R1.M
1 000
PS 3500.R1.M
2 000
PS 4500.R1.M
2 000
PS 6100.R1.M
5 000