Mode:
Continuous Monitor, Passive Diffusion
Principle of Operation:
Pulse Mode Ionization Chamber
Minimum Sensitivity:
0.27 cpm/pCi/L (minimum 64 alpha counts per hour @ 4 pCi/L)
Units:
picoCuries per liter (pCi/L)
Dynamic Range:
0.5 to 150 pCi/L
Linearity:
+/- 10% Best Straight Line
Data Storage (Nonvolatile):
Up to 240 hourly readings, wrap on over-run.
Power:
120VAC 60Hz Transformer, 8 days rechargeable battery operation.
Indicators:
Green AC Power LED, Red Testing LED
Key Switch:
3 Position, Removable Key [ Off ] [ Keypad On ] [ Measure On ]
Pushbuttons:
Momentary [ Print ] [ Hold to Clear ]
Communications:
RS-232 port for printing or computer uploads
Tamper Protection:
Tilt switch, AC power fail detection, key switch lock-out.
Operating Range:
Temperature 50
̊
F – 90
̊
F Humidity 0-80%, non-condensing
Size:
Height
7 inches (includes Carrying Handle)
Width
8 inches
Depth
4 ½ inches
Weight
2 ½ pounds
Color:
Green/Black
Mounting:
(2) ¼”x 20 Threaded Tripod Mounts
RS300 Instrument Specifications
Manufactured by
RadonAway
3 Saber Way
Ward Hill, MA 01835
Technical Support, Calibration
and Repair Service.
800-767-3703
M-F 8:30am - 5:30pm ET