4

5.0 OBTAINING A READING FROM THE RS300:

5.1 Any time during or after a test turn the keyswitch to [Keypad On] position.

Note: To continue testing at anytime when the keyswitch has been turned to the

[ Keypad On ] or the [ Off ] position, simply turn the keyswitch to [ Measure On ]

position. Note that the unit will “beep” to signify continued testing.

Note: When a test is continued by going from [ Keypad On ] to [ Measure On ] the

hour that you were testing in starts over again. Also note that if you try to look at

the readings before the first hour is up the PC or printer will read: “No Data”.

6.0 SETTING UP THE OPTIONAL PRINTER :

6.1 Plug the AC/DC adapter (the one marked “Printer Power Supply”) into any 120VAC

outlet. Plug the power jack from the adapter into the back of the printer.

6.2 Connect the D-Sub cable supplied with the unit to the printer and the RadStar.

6.3 Turn the printer on via the rocker switch. The red LED on the printer should come on.

7.0 OBTAINING A TEST REPORT PRINTOUT:

7.1 After completing a test turn the keyswitch to [ Keypad On ] position.

7.2 Press and hold down [ Print ] button until unit beeps. The unit will print the test

report. See Fig.A (p.6) for an explanation of the printout.

Note: If you try to get a printout within the first hour of testing, the printer will

display “No Data to Print”.

7.3 To obtain multiple printouts, simply press the [ Print ] button again.

Содержание RadStar RS300

Страница 1: ...tor Light Power Jack Printer Port 12 or 24 hour start time delay option Internal memory to retain data Print Button Clear Button Key Switch INO28 Rev D Accurate measurement in a wide range of temperatures and relative humidity Nominal sensitivity 16 cph pCi L ...

Страница 2: ...RADON GAS 5 0 OBTAINING A READING FROM THE RS300 6 0 SETTING UP THE OPTIONAL PRINTER 7 0 OBTAINING A TEST REPORT PRINTOUT 8 0 DOWNLOADING DATA FROM THE RADSTAR 9 0 TRANSPORTING THE RADSTAR TO A PC OR PRINTER FIGURE A EXAMPLE RS300 TEST RESULT PRINTOUT Table of Contents 3 3 3 3 4 4 4 5 5 6 ...

Страница 3: ...on This tells you that the NiCad batteries are charging Power cord should be left plugged in at all times during a radon test to ensure that the batteries remain charged If there is no power available such as in new construction the unit may be run for a typical 48hr test on the batteries alone as long as the unit has been charging with the power adapter over night 4 0 MONITORING FOR RADON GAS 4 1...

Страница 4: ...TTING UP THE OPTIONAL PRINTER 6 1 Plug the AC DC adapter the one marked Printer Power Supply into any 120VAC outlet Plug the power jack from the adapter into the back of the printer 6 2 Connect the D Sub cable supplied with the unit to the printer and the RadStar 6 3 Turn the printer on via the rocker switch The red LED on the printer should come on 7 0 OBTAINING A TEST REPORT PRINTOUT 7 1 After c...

Страница 5: ... Set Delay menu will appear No Delay 12 Hour Delay or 24 Hour Delay must be selected Note To meet EPA protocols minimum test durations are 48 60 and 72 hours for No Delay 12 Hour Delay or 24 Hour Delay respectively On the RadStar press the Print button once The data will download to the PC Note Clicking Download Data on the PC after pressing Print on the RadStar Unit will result in incomplete data...

Страница 6: ...FIGURE A EXAMPLE RS300 TEST RESULT PRINTOUT 6 ...

Страница 7: ...Manufactured by RadonAway 3 Saber Way Ward Hill MA 01835 ...

Страница 8: ... Testing LED Key Switch 3 Position Removable Key Off Keypad On Measure On Pushbuttons Momentary Print Hold to Clear Communications RS 232 port for printing or computer uploads Tamper Protection Tilt switch AC power fail detection key switch lock out Operating Range Temperature 50 F 90 F Humidity 0 80 non condensing Size Height 7 inches includes Carrying Handle Width 8 inches Depth 4 inches Weight ...

Отзывы: