1
Using eCAL to retrieve the original factory
calibration certificate
The original factory calibration results for the locator can be retrieved from
the unit. Each time the is calibrated either at Radiodetection or an approved
Radiodetection service centre, the calibration results are stored within the locator. To
retrieve the results and print a certificate, carry out the following:
Note:
You do not need to purchase an eCAL validation key to retrieve the original
factory calibration certificate.
Register your locator by going to:
www.radiodetection.com/extendedwarranty
. See page for more details.
Download Centros Manager by going to:
www.radiodetection.com/centrosmanager
. See page 1 for details.
Connect the via the USB connector inside the battery compartment
to a suitable USB port on a PC or laptop.
Switch on the (no segments will be lit but the backlight will be on).
Open Centros Manager and click on Locator eCAL Validation.
Click on
Get Original Calibration Data
and a message box will automatically pop
up displaying the progress.
In less than 3 minutes the original calibration certificate will be available to view
or print, locate the serial number of the within the Unit Manager window
and expand the contents. Expand Calibration History and click on each date.
When you click on each date, the type of certificate, whether an eCAL validation or
original factory calibration, together with date and serial number will be displayed.
By clicking on this window, the certificate will be displayed and you can choose to
print this certificate.