Ref No.
:
082012
Normal Body Temperature
Normal body temperature is a range that will fluctuate throughout the day. The normal range for adult body temperature is typically considered to
be 36.1 to 37.8° C (97 to 100°F). The body temperature varies by age, person, gender, time of day, position of body, and is usually highest in the
evening. It can be affected by activity, emotion, clothing, medications, ambient conditions, hormones and other factors. Normal temperature will
also be different depending on the place on the body at which the temperature is taken, with rectal or ear temperature readings generally being
higher than oral temperature readings, and armpit temperature readings generally lower than oral readings.
Temperature variations affected by age:
Clinical repeatability: 0.18
℃
(< 1 year old), 0.17
℃
(1~5years old), 0.15
℃
(>5years old)
Important Notes
Trouble shooting
Warranty
Warranty :12 months
Manufacture Date
:
as the serial number( please open the battery cover, it is shown on the inside of the device.)
Ex.SN:E912A000001, the first “E” is External, the second number “9” is the manufacture year 2009, the third and the
fourth number “12” is the manufacture month, the others is the serial number.
Note: The thermometer is calibrated at the time of manufacture. If at any time you question the accuracy of
temperature measurements, please contact the dealers or nearest service address.
Please read the instructions for use
BF type applied part
Portable and mobile RF communication equipment can affect medical electrical equipment such as this product.
This device complies with
CISPR 11 group 1, Class B. During measurement it is recommended to keep at least 1.5 meter separation between this device and the
surrounding electrical devices. If interference occurs, please increase the separation further or turn off the electrical equipment in the
vicinity. For detailed description regarding the EMC precautions you need to take, please contact our authorized representative or us on
address as given below.
Cleaning
And Storage
T
he probe is the most delicate part of the thermometer. Use with care when cleaning the
lens to avoid damage.
** Replace the probe cover after each use to ensure an accurate reading and
avoid cross contamination
.
a. Storage temperature range: It should be stored at room temperature
between -20~+50°C
,
RH
≦
85%
b.
Transportation temperature shall be less than 70°C, RH
≦
95%
c. Keep the unit dry and away from any liquids and direct sunlight.
d. The Probe should not be submerged into liquids
*
* If device is accidentally used without probe cover, clean the probe as follows:
a. After the measurement, please use the cotton swab with the Alcohol (70% concentration)
to clean the lens( on the inside of the probe).
b. Allow the probe to fully dry for at least 1 minute.
Note:
Please check if the device is damaged once it drops. If you can’t make sure of it, please
send the complete device to your local dealer for recalibration.
Holding the thermometer too long may cause a higher ambient temperature reading of
the probe. This could make the body temperature measurement lower than usual.
Battery
Replacement
This device is supplied with one lithium cell CR2032x1.
1. Open the battery cover: Insert a pointed object into battery cover pick hole. At the same
time, use thumb to push battery cover out.
2. Hold the device and flip the battery out with a small screwdriver.
3. Insert the new battery under the metal hook on the left side
○
1
and press the right side
○
2
of the battery down until you hear a “Click”.
4. Replace the battery cover
Keep the battery away from children.
The positive (+) side up and the negative (-) side pointed down.
Error Message
Problem
Solution
The probe cover has not been installed well
Please check the setting of the probe cover again.
(Refer to the section of “Place the Probe Cover” )
Error 5~9, the system is not functioning properly.
Unload the battery, wait for 1 minute and repower it. If the
message reappears, contact the retailer for service.
Measurement before device stabilization.
Wait until all the icons stops flashing.
The device showing a rapid ambient temperature change.
The ambient temperature is not within the range between
10
o
C and 40
o
C (50
o
F ~104
o
F).
Allow the thermometer to rest in a room for at least 30
minutes at room temperature: 10
o
C and 40
o
C (50
o
F ~104
o
F).
Temperature taken is higher than +42.2
o
C (108°F)
Check the integrity of the probe cover and take a new
temperature measurement.
Temperature taken is lower than +34
o
C (93.2°F)
Make sure the probe cover is clean and take a new
temperature measurement.
Device cannot be powered on to the ready stage.
Change with a new battery.
Radiant Innovation Inc. Http://www.radiantek.com.tw
Add: 1F, No.3, Industrial East 9
th
Road, Science-Based Industrial Park, HsinChu, Taiwan 300.
Medical Technology Promedt Consulting GmbH
Add: Altenhofstrasse 80, D-66386 St. Ingbert, Germany
A
G
E
T
e
m
p
.
Young
Old
High
Low
0 ~ 2 years:
36.4°C (97.5°F) ~ 38.0°C (100.4°F)
3 ~ 10 years:
36.1°C (97.0°F) ~ 37.8°C (100.0°F)
11 ~ 65 years:
35.9
°°°°
C (96.6°F) ~ 37.6
°°°°
C (99.7°F)
> 65 years:
35.8
°°°°
C (96.4°F) ~ 37.5
°°°°
C (99.5°F)