Symbol Descriptions 

 

The CE mark and Notified Body 
Registration Numbers, the 
requirement of Annex II from Medical 
Device Directive 93/42/EEC are met.

 

 

Please do not dispose of the product in the 
household waste at the end of its useful life. 
Disposal can take place at appropriate 
collection points provided in your country. 

 

Authorized 
representative in the 
European community 

 

Caution 

 

Follow instructions for use 

 

Manufacturer 

 

BF type applied part 

 

To protect the environment, dispose of empty 
batteries at your retail store or at appropriate 
collection sites according to national or local 
regulations. 

 

Paper Recycling 

IP22

 

Classification for water ingress and 

particulate matter. 

 

 

 

 

Manufacturer’s declaration-electromagnetic emissions 

The TH09F is intended for use in the electromagnetic environment (for home healthcare) specified below.The customer or the user of the TH09F should 
assure that it is used in such an environment. 

Emissions test 

Compliance

 

Electromagnetic environment 

– 

guidance 

(for home healthcare environment)

 

RF emissions 
CISPR 11 

Group 1 

The TH09F uses RF energy only for its internal function. Therefore, its RF emissions are very 
low and are not likely to cause any interference in nearby electronic equipment.   

RF emissions 
CISPR 11 

Class B 

The TH09F is suitable for use in all establishments, including domestic establishments and 
those directly connected to the public low-voltage power supply network that supplies buildings 
used for domestic purposes. 

Manufacturer’s declaration – electromagnetic immunity 

The TH09F is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the TH09F should 
assure that it is used in such an environment. 

Immunity test 

IEC 60601 test level

 

Compliance level

 

Electromagnetic environment 

– 

guidance 

(for home healthcare environment)

 

Electrostatic discharge(ESD) 

IEC 61000-4-2 

Contact:±8 kV   

Air±2 kV,±4 kV,±8 

kV,±15 kV 

Contact:±8 kV   

Air±2 kV,±4 kV,±8 

kV,±15 Kv 

Floors should be wood, concrete or ceramic tile. If floors are covered 

with synthetic material, the relative humidity should be at least 30% 

Power frequency(50, 60 Hz) 

magnetic field IEC 61000-4-8 

30 A/m 

50 Hz or 60 Hz 

30 A/m 

50 Hz and 60 Hz 

The TH09F power frequency magnetic fields should be at levels 

characteristic of a typical location in a typical home healthcare 

environment. 

Manufacturer’s declaration – electromagnetic immunity 

The TH09F is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the TH09F should 
assure that it is used in such an environment. 

Immunity test 

IEC 60601 test level

 

Compliance 

level

 

Electromagnetic environment 

– 

guidance 

(for home healthcare environment)

 

Radiated RF   

IEC 61000-4-3 

10 V/m 

80 MHz 

– 2,7 GHz   

80 % AM at 1 kHz 

10 V/m 

80 MHz 

– 2,7 

GHz   

80 % AM at 1 

kHz   

Recommended separation distance: 

d = 1,2 

√P

 

d = 1,2 

√P

   

80MHz to 800 MHz 

d = 2,3 

√P

    800MHz to 2,7 GHz   

Where

 P

 is the maximum output power rating of the transmitter in watts (W) according to 

the transmitter manufacturer and 

is the recommended separation distance in metres 

(m). 
 
Field strengths from fixed RF transmitters, as determined by an electromagnetic site 
survey, 

a

 should be less than the compliance level in each frequency range.

 
Interference may occur in the vicinity of equipment marked with the following 
symbol: 

NOTE1    At 80 MHz and 800 MHz, the higher frequency range applies. 

NOTE2    These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects 

and people. 

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM 
radio  broadcast  and  TV  broadcast  cannot  be  predicted  theoretically  with  accuracy.  To  assess  the  electromagnetic  environment  due  to  fixed  RF 
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the TH09F is used exceeds the 
applicable  RF  compliance  level  above,  the  TH09F  should  be  observed  to  verify  normal  operation.  If  abnormal  performance  is  observed,  additional 
measures may be necessary, such as re-orienting or relocating the TH09F. 

Содержание TH09F

Страница 1: ...m from the central forehead Fig 1 and press the START button to get the forehead measurement The time consuming for measurement might be 1 second After two short beeps are heard the icon stops flashing to be ready for next measurement Figure 1 Points for attention a Forehead temperature is displayed in oral mode This mode converts the forehead temperature to display its oral equivalent value b Bef...

Страница 2: ...it for Er1 to disappear The ambient temperature is not within the range between 10o C and 40o C Allow the thermometer to rest in a room for at least 15 minutes at room temperature 10o C and 40o C Temperature taken is higher than 43o C Please select the target within specifications If a malfunction still exists please contact the nearest retailer Temperature taken is lower than 28o C Device cannot ...

Страница 3: ...ic tile If floors are covered with synthetic material the relative humidity should be at least 30 Power frequency 50 60 Hz magnetic field IEC 61000 4 8 30 A m 50 Hz or 60 Hz 30 A m 50 Hz and 60 Hz The TH09F power frequency magnetic fields should be at levels characteristic of a typical location in a typical home healthcare environment Manufacturer s declaration electromagnetic immunity The TH09F i...

Страница 4: ...er s declaration electromagnetic immunity Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment The TH09F is intended for use in the electromagnetic environment for home healthcare specified below The customer or the user of the TH09F should assure that it is used in such an environment Test frequency MHz Band a MHz Service a Modulation b Maximum power W Distance ...

Отзывы: