RD-30 Operations Manual
Operations Overview
20
4.9 Analog Sense Input
The optional Analog Sense functionality allows the RD-30 to effectively test transducers as well
as first generation electronic meters that have an analog output for calibration. The RD-30 will not
only test these instruments, but it will also automatically calculate and display the error.
4.8 Pulse Outputs
The RD-30 Pulse Outputs consist of BNC connectors labeled 1, 2, 3. These outputs are open
collector. The factory default setting is Output 1 = Total Watthours, Output 2 = Total VARhours,
Output 3 = Total VAhours. However all three outputs are programmable by the customer for per
phase energy or total energy.