Dynalyzer – Digital Display/Printer
OPERATION MANUAL
24
Output Port
RS232 Port Settings
This RS232 port is set to transmit data using the following parameters:
•
2400
baud
•
No parity checking
•
8 data bits
•
1 stop bit
The receiving equipment should be set to the same parameters.
The Dynalyzer will output data on the RS232 port with the exact same format as the printer data. Only a
CARRIAGE RETURN (ODh) is sent at the end of each line, so the receiving equipment may have to be set up
to add a LINE FEED if necessary. The Dynalyzer only outputs data to the RS232 port. There are not provisions
to receive data or commands from external equipment.