background image

 

 

 

 

Important: From this instant and up to returning back to the Main Menu, all control shall 
be  made  from  the  keyboard  only.  This  is  made  specifically  for  convenient  use  of  the 
computer in the field condition. In the Data Acquisition dialog box, use keys 

 and 

 to 

select Setup and press Enter or Space. 
 

5. Enter 

Setup

 menu. Once 

Setup

 is activated, the screen shows 

ZOND SYSTEM 12e SETUP

   

dialog box and set all the parameters the same as on the picture below. Menu options are selected 
using cursor keys and Enter or Space. At the first run signal may be not seen or located not in the 
same position as  on picture. To set  correct  position of signal  enter 

Pulse  delay

  menu item  and 

press “

A

” key on keyboard to start automatic adjustment. Just adjustment is finished press Enter 

and then 

Close 

setup menu. 

 

 

 

Fig. 5. Setup menu. 

 
 
Important:  Zond-12e  500A  GPR  has  built-in  deactivating  mechanism,  i.e.  transmitter  is 
switched  ON  only  when  operator  enters  SETUP  menu  item  or  START  data  acquisition. 
Transmitter  switches  OFF  immediately  after  operator  STOP  data  acquisition  or  leaves 
SETUP menu item. 
 

2.3. Data acquisition.  

 

Before to start survey please put Ethernet cable into groove on housing case wall and then close 
housing GPR case as it is shown on Fig.6. If Ethernet cable is too short it is possible to add one 
more cable using the coupler included in accessories kit. 
 
 
 

 

Содержание Zond-12e

Страница 1: ...Zond 12e 500A Ground Penetrating Radar USER S MANUAL Riga 2019 Darzauglu Str 1 105 Riga LV 1012 Latvia Phone Fax 371 67141041 E mail radsys radsys lv Web www radsys lv...

Страница 2: ...consult with your network administrator before changing IP addresses Set Subnet mask as 255 255 255 0 The computer performs the following functions it controls operation modes of the GPR and receives...

Страница 3: ...3 Fig 1 Zond 12e 500A Ground Penetrating Radar in shockproof case Fig 2 View of shock proof case control panel of Zond 12e 500A GPR...

Страница 4: ...the User s Manual for Prism2 program Once the installation is completed you must perform the actions prescribed by the Manual in paragraphs 5 1 and 5 2 Configuring the computer to connect with Zond 12...

Страница 5: ...he same position as on picture To set correct position of signal enter Pulse delay menu item and press A key on keyboard to start automatic adjustment Just adjustment is finished press Enter and then...

Страница 6: ...Zond 12e 500A GPR Just SETUP menu is closed button START is active Press Enter or Space key Display shows data as on picture below To stop data acquisition press Esc key or Enter Space when STOP butto...

Страница 7: ...gher than 40 C it is not recommended to leave the instrument in place where it can be affected by direct sun stares 4 TRANSPORTATION RULES 4 1 In case of observing of rules of packaging of the instrum...

Страница 8: ...ting to Electromagnetic Compatibility and Radio Spectrum Matters 99 5 EC Short range devices Ground and Wall Probing Radars applications Low voltage Directive 73 23 EEC and the Amendment Directive 93...

Страница 9: ...uld void the user s authority to operate the equipment Operation of this device is restricted to law enforcement fire and rescue officials scientific research institutes commercial mining companies co...

Страница 10: ...Domestic United States users The Federal Communications Commission FCC requires that all users of GPR who purchased antennas after July 15th 2002 register their equipment and areas of operation If you...

Страница 11: ...8 2 R glement d Industrie Canada Fran ais Ce dispositif est conforme la norme CNR 220 d Industrie Canada applicable aux appareils radio exempts de licence Son fonctionnement est sujet aux deux condit...

Страница 12: ...12 9 OUR ADDRESS Radar Systems Inc Darzauglu Street 1 105 Riga LV 1012 LATVIA Phone 371 67141041 E mail radsys radsys lv...

Отзывы: