1260-100X User Manual
1260-100X Specifications A-2
Racal Instruments
©
2005
Power and
Cooling
Power Consumption
+5V:
1.0A
(without plug-ins)
+12V:
0A
+24V:
0A
Power Available
+5V:
2.5A
(per plug-in)
+12V:
1.0A
+24V:
1.0A
Cooling (worst-case
Airflow:
3.0
liters/sec
plug-in configuration)
Backpressure:
0.7 mm H
2
0
Environmental
Temperature
Operating:
0
°
C to +55
°
C
Non-operating: -40
°
C to +70
°
C
Relative Humidity
95% RH non-condensing
Altitude
Operating:
10,000
ft.
Non-operating: 15,000
ft.
Vibration
0.013 in. P-P, 5 Hz to 55 Hz; meets
MIL-T-28800C Type III, Class 5,
Style F
Shock
30g, 11 ms, 1/2 sine wave
Bench Handling
4 inch drop at 45
°
EMC and Safety
Emissions
EN61326: 1997+A1:1998, Class A
Susceptibility, Safety
EN61010-1:1993+A2:1995
Reliability
Mean Time Between
646,921hours, MIL-HBK-217,
Failures (MTBF)
ground-benign, 25
o
C
Mean Time To Repair
Replace plug-in: 5 min.
(MTTR)
Replace other:
<30 min.
*Measured from start of VXIbus cycle until relay coil is fully energized.