1260-100X User Manual
Maintenance 4-2
Racal Instruments
©
2005
Software
Required
•
Microsoft Windows 95, 98, ME, NT, 2000, and XP
•
VXI Resource Manager (RESMAN) by National
Instruments
•
VXI Interactive Control (VIC) by National Instruments
Procedure
1. Ensure that an Option-01T is installed in the
carrier adjacent to the 1260-100X carrier as
outlined in the
Option-01T
section of
Chapter 2:
Getting Started
.
2. Set the Option-01T logical address switch to 3.
3. Set the
1260-100X
module address switch to the
1-6 position.
4. Install the carrier into an empty pair of slots in a
VXIbus chassis. Remove all other modules from
the chassis except the slot 0 controller and the
carrier with an Option-01T.
5. Insert one or more plug-ins into the 1260-100X
Carrier.
6. Turn the chassis power on. Verify that the Fail
LED on the carrier with an Option 01T illuminates,
then extinguishes within five seconds.
7. Run the National Instruments Resource Manager
program (RESMAN). From the RESMAN output
log file, verify the following:
8. The Option-01T is recognized as DEVICE_3.
9. The self-test for DEVICE_3 passed.
10.
16K-bytes of memory are allocated for
DEVICE_3.
11. Start the VXI Interactive Control (VIC) program
(by National Instruments).