1260-145 User Manual
Module Specification 1-16
Racal Instruments
2003
module. If five additional modules are likewise loaded, then the
overall carrier dissipation is approximately 80W, which is within the
cooling available in most commercial VXIbus chassis. Apart from
the thermal issues, however, it is important to remember that a
fully enabled 1260-145E card would draw over 2.7A of current from
the +5V supply. If this is multiplied by six for the remaining slots, a
load of 16A would burden the +5V supply, exceeding the
specifications for the 1260-100 carrier. In reality, the worse-case
situation is not likely, since the 1260-145E is a matrix switch where
typically only one relay is engaged at a time and the power
dissipation falls considerably below that calculated in the example
above. For practical purposes, as the calculations can illustrate for
normal operation, the 1260-145E contributes a negligible amount
of thermal load on a chassis.
Introduction –
1260-145F
(1 – 8x16)
The 1260-145F is a plug-in matrix switch module developed for the
Racal Instruments 1260-100 Adapt-a-Switch Carrier. This switch
provides a 8x16 matrix with a maximum switching voltage of
125VAC @ 0.3A or 30 VDC @ 1A.
The 1260-145F includes the following features:
•
Standard Adapt-a-Switch
plug-in design, providing for ease
of replacement
•
Data-Driven embedded descriptor, allowing immediate use with
any Option-01T switch controller, regardless of firmware
revision level.
Содержание 1260 VXI
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 10: ...1260 145 User Manual iv This page was left intentionally blank...
Страница 34: ...1260 145 User Manual Module Specification 1 24 Racal Instruments 2003...
Страница 60: ...1260 145 User Manual Optional Harness Assemblies 4 4 Racal Instruments 2003...