Publication No. 980673-055 Rev. A 1260-22 User Manual
EADS North America Test and Services Module Operation 3-3
Multiple individual channels may be specified using the following
channel descriptor syntax:
(@ <module address> ( <chan1> , <chan2> , .
. ., <chanN> ))
A range of channels may be specified using the following channel
descriptor syntax:
(@ <module address> ( <first channel> :
<last channel> ))
The following examples illustrate the use of the channel
descriptors for the 1260-22:
OPEN (@8(0))
Open channel 0 on the 1260-22 with module
address
8
CLOSE (@8(0,7))
Close channels 0 and 7 on the 1260-22 with
module
address
8.
CLOSE (@2(7:12))
Close channels 7 through 12 inclusive on
the 1260-22 with module address 2
Reply To The
MOD:LIST?
Command
The 1260-01T returns a reply to the MOD:LIST? command. This
reply is unique for each different 1260 series switch module. The
syntax for the reply is:
<module address> : <module-specific identification string>
The <module-specific identification string> for the 1260-22 is:
1260-22 40-CHANNEL SPST 20A SWITCH MODULE
So, for a 1260-22 whose <module address> is set to 8, the reply to
this query would be:
8 : 1260-22 40-CHANNEL SPST 20A SWITCH MODULE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com