1260-100/101 Adapt-a-Switch Carrier Module User Manual
C-1
Appendix C
TROUBLESHOOTING
If you have difficulty in using the 1260-100/101 Adapt-a-Switch Carrier or its plug-ins, refer to the
following troubleshooting table. If you are still unable to resolve the problem, contact one of the
Racal Customer Support offices listed in
Chapter 5: Support Offices
.
Troubleshooting Table
PROBLEM
POSSIBLE CAUSES
WHAT TO DO
FAIL indicator does not come on
when chassis power is turned on.
•
Option-01T not installed
correctly.
•
Fail LED cable not connected
properly to Option-01T.
Remember: Fail LED is operational
only on module containing
Option-01T.
•
Review
Installing the Option-01T
in
Chapter 2: Getting Started
•
Verify that cable is correctly
connected. See
Installing the
Option-01T
in
Chapter 2: Getting
Started
.
FAIL indicator stays on after self-
test is completed.
•
Option-01T defective.
•
Contact Racal Customer Support
(see
Chapter 5: Customer
Support
).
No response from plug-in via
VXIbus.
•
Option-01T not installed, or
installed in incorrect module.
•
Module addresses conflicting.
•
Plug-in not fully seated.
•
Communication problem with
Option-01T, 1260-100/101
Carrier, or plug-in.
•
Option-01T must be installed in left-
most carrier or 1260-Series module
in group. See
Option-01T
in
Chapter 2: Getting Started
.
•
Ensure that each plug-in and
1260-Series module has its own
unique module address. See
Module Address Switch
in
Chapter 2: Getting Started
.
•
Ensure that plug-in is firmly
engaged.
•
See special troubleshooting
procedure below, entitled
“Communications Troubleshooting”
Содержание 1260-100
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 7: ......
Страница 8: ...This page was left intentionally blank...
Страница 12: ...1260 100 101 Adapt a Switch Carrier Module User Manual iv This page was left intentionally blank...
Страница 53: ...1260 100 101 Adapt a Switch Carrier Module User Manual C 4 This page was left intentionally blank...