39
support.R82.org
Electromagnetic compatibility (EMC) environ-
ment
This equipment is for use in home healthcare
facility environment.
This equipment is not likely to cause any
harmful interference in nearby equipment;
however, it can be impaired by electromagnetic
(EM) interference emitted by other equipment,
which can lead to malfunctions. Do not use
this equipment where there is a risk that other
electrical and/or electronic equipment can
interfere with this equipment.
This equipment complies with the following
standards
IEC 60601-1-2:2014 Medical electrical
equipment - Part 1-2: General requirements
for basic safety and essential performance
- Collateral Standard: Electromagnetic
disturbances - Requirements and tests (EN
60601-1-2:2015)
EN 12182:2012, Clause 7, Assistive products for
persons with disability – General requirements
and test methods
EMC - WARNINGS
Use of this equipment adjacent to or stacked with
other equipment should be avoided because it
could result in improper operation. If such use
is necessary, this equipment and the other
equipment should be observed to verify that they
are operating normally.
Use of accessories, transducers and cables
other than those specified or provided by
R82 could result in increased EM emission or
decreased EM immunity of this equipment and
result in improper operation. Several power
plugs are supplied with this equipment. Please
use the correct power plug suitable for your
country. Please chose among the supplied power
plug adaptor units: English, Australian, EU, US/
Japanese plugs.
Portable radio frequency (RF) communication
equipment (including peripherals such as
antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to
any part of the Wombat Living, including cables
specified by R82. Otherwise, degradation of the
performance of this equipment could result.
Testing for EM disturbances (EM emission and EM immunity)
Test modes during emission tests:
1.
Movement mode (battery). Actuator running in/out
2.
Charging mode, actuator not running.
EM emission standards
Classification/levels
Conducted and radiated RF emission CISPR
11
Group 1 (the product uses RF energy only for its
internal function)
Class B (the product is suitable for use
in all establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes)
Harmonic distortion
IEC 61000-3-2
Class A.
No testing performed, as the rated power for
this equipment is less than 75 W. No limits are
specified for equipment with a rated power of 75
W or less.
Voltage fluctuations and flicker
IEC 61000-3-3
dmax ≤ 4 % (without additional conditions)
EN
Содержание Wombat Living 1
Страница 3: ...3 support R82 org WOMBAT LIVING 1 2 WOMBAT LIVING 3 01 70 kg 154 lb 45 kg 99 lb ...
Страница 4: ...4 support R82 org ...
Страница 5: ...5 support R82 org 1 3 A 2 ...
Страница 6: ...6 support R82 org 1 5 3 B 2 5 mm 2 5 Nm 4 ...
Страница 7: ...7 support R82 org 1 3 C 2 ...
Страница 8: ...8 support R82 org 2 3 4 Size 1 2 1 5 6 ...
Страница 9: ...9 support R82 org 5 mm 10 mm 2 3 4 Size 3 1 5 5 mm 10 mm 6 ...
Страница 10: ...10 support R82 org 2 3 4 1 ...
Страница 11: ...11 support R82 org 05 05 2 3 5 4 1 6 ...
Страница 12: ...12 support R82 org 05 2 3 5 6 4 1 ...
Страница 13: ...13 support R82 org 05 2 3 5 4 1 ...
Страница 14: ...14 support R82 org 2 3 5 6 4 1 ...
Страница 15: ...15 support R82 org 2 3 4 1 5 ...
Страница 16: ...16 support R82 org 2 3 4 1 ...
Страница 17: ...17 support R82 org 2 3 4 1 ...
Страница 18: ...18 support R82 org 2 3 4 1 5 ...
Страница 19: ...19 support R82 org 2 3 4 Size 3 1 5 ...
Страница 20: ...20 support R82 org 2 3 5 4 1 6 ...
Страница 21: ...21 support R82 org 2 3 5 4 1 ...
Страница 22: ...22 support R82 org 2 3 1 ...
Страница 23: ...23 support R82 org 1 2 3 4 ...
Страница 24: ...24 support R82 org 4 4 3 5 4 4 4 1 2 6 ...
Страница 25: ...25 support R82 org BA 18 LA 28 HB 40 ...
Страница 144: ...support R82 org ...