Quicktest KEE M-509GM Скачать руководство пользователя страница 7

FREQUENTLY ASKED QUESTIONS 

 
How long does the probe last? 

 

Approximately, 1000 tests - but only if the cap is replaced firmly after every 
use. In practice, it will give 3 to 4 months of heavy use in a gold-buying shop, 
or up to a year if used just occasionally. After a year, it should be replaced. 
 
If the cap is left off for a few hours (e.g. overnight) it 

will

 dry out.  

We strongly recommend that you keep a spare probe in stock. 
 

How long does the battery last? 

 

A few months. If you leave it on (continuously) it will last a week or two. 
 

How often need the probe be calibrated? 

 

Keep the 18ct sample on the plate, touch it every few tests to check the meter 
registers “18kt”.  At first, you will find it stays in calibration for weeks.  As the 
fluid in the probe gets used up you will have to calibrate it more often.  
Eventually you will have to calibrate it every few tests, this is a sign that the 
probe refill will soon need replacing. 
 

What if the meter doesn’t move to the far right upon start up?  

 

Change the battery.  

 

What if it won’t calibrate to the middle of “18kt”?   

 

First, change the battery

. If that doesn’t work, change the probe refill. 

 

Can more than one item be placed on the test plate at once?

 

 

Yes. For easy checking you may like to keep the calibration sample on the 
test plate all the time. For easy testing you may like to place a few items on 
the test plate at once then test them one after the other.  
 

Why is the dial m

arked “Krt” and not “ct”?

 

 

Because it

’s American.  They use the abbreviation “K” (or “Kt”) for karat 

instead of “ct” for Carat (and the manufacturer has mistakenly printed “krt”).  
Any reading in the 10krt area indicates 9ct, whether it is lower or higher than 
the middle of the 10krt (yellow) area. 

 

Содержание KEE M-509GM

Страница 1: ...LECTRONIC GOLD TESTER MODEL M 509GM www quicktest co uk All equipment must be used in conjunction with the user s skill knowledge and experience Under no circumstances shall Quicktest be liable for direct or indirect loss sustained in connection with testing equipment It is your responsibility to regularly check readings against known samples ...

Страница 2: ... samples on the test plate and then touch the top of the sample with the probe tip Is it necessary to clean the side of the sample that is resting on the test plate If it is covered with a layer of dirt then yes otherwise no there is no need to clean it you may if you wish but it s not necessary The item must make firm contact with the plate if it does not have a flat surface that makes firm conta...

Страница 3: ...use either F Black socket on tester G Abrasive tip H Protective snap cap on probe I Probe tip J Replaceable part of probe K The end of the probe with the red socket L The red socket on the end of the probe M The red plug on the red lead they are both the same N The black plug on the black lead O Crocodile clip ...

Страница 4: ...rea Remove the probe tip carry out the test again this time the needle will settle more quickly to be certain it calibrated correctly If you are not certain carry out the test a third time For standard testing of jewellery this method of calibration is accurate enough For testing 20ct to 24ct see TESTING HIGH CARAT ACCURATELY on the next page Now that the tester is calibrated test your unknown sam...

Страница 5: ...ld be about 22ct you may wish to know if it is slightly more or slightly less than 22ct Calibrate the tester on your 18ct calibration sample then calibrate it on a 22ct sample You might easily calibrate them both to the middle of the 18ct and 22ct area of the dial or you might have to compromise by setting one or both of the samples fractionally higher or lower This is fiddly but it is the only wa...

Страница 6: ...W ON THE DIAL It takes one AA battery Use a good quality alkaline or Lithium battery The outer sleeve is made of a soft rubber it simply pushes off as shown in the photographs below Then remove the two screws open the tester replace the battery ...

Страница 7: ...e gets used up you will have to calibrate it more often Eventually you will have to calibrate it every few tests this is a sign that the probe refill will soon need replacing What if the meter doesn t move to the far right upon start up Change the battery What if it won t calibrate to the middle of 18kt First change the battery If that doesn t work change the probe refill Can more than one item be...

Отзывы: