.Calibrating the Temperature 

The soldering iron should be recalibrated after changing the iron.   

Tool: QUICK 191/192 thermometer 

Method:   

1. Press and hold “1” and “3” buttons at the same time to enter into calibration menu. 

2. The LCD display is flashing, press “+” and “-” button to enter value. 

3. When the value is the same as the thermometer, press “STORE” to calibrate it. 

 

 

.Tip Care and Maintenance 

Select Tip 

1.

 

Select  a  tip  that  maximizes  contact  area  between  the  tip  and  solder  joint. 

Maximizing  contact  area  gives  the  most  efficient  heat  transfer,  allowing 

operators to produce high quality solder joints quickly.   

2.

 

Select  a  tip  that  allows  good  access  to  the  solder  joint.  Shorter  tip  lengths 

allow  more  precise  control.  Longer  or  angled  may  be  needed  for  soldering 

densely populated boards. 

 

Содержание TS1300

Страница 1: ...nstruction Manual Thank you for purchasing our Lead Free Soldering Station It is designed for lead free soldering Please read this manual before operating the unit Store this manual in a safe easily a...

Страница 2: ...Parameters Setting 8 6 1 Enter into SET menu 8 6 2 Temperature setting 9 6 3 Alarm Temperature Setting 9 6 4 Alarm Temperature Setting 10 6 5 Sleep time setting 10 6 6 Close time setting 11 6 7 Temper...

Страница 3: ...ace parts or install tips turn the power off and allow the unit to cool to room temperature To prevent damage to the unit and ensure a safe working environment be sure to comply with the following pre...

Страница 4: ...t When your hands are wet don t use and disconnect the unit or to pull the supply cord The soldering process will produce smoke so make sure the area is well ventilated While using the unit don t do a...

Страница 5: ...ls for lead free soldering 1 LCD display 2 Touch switch Power soft switching 3 Mounting positioning edge 4 Handle socket 5 Power switch Hard switching 6 Power supply socket 7 ESD socket 8 RS485 port 9...

Страница 6: ...re Range 50 450 122 842 Operation ambient Temperature Max Max 40 Tip to Ground Resistance 2 Tip to Ground Potential 2 mV Dimension W H D 190 190 134mm Weight 2 7 Kg Ensure that your power supply data...

Страница 7: ...en the cleaning sponge with water and then squeeze it dry 2 Place it in groove of the iron holder base 3 Add a little water to holder base The sponge will absorb water to keep the sponge around it wet...

Страница 8: ...s and hold POWER button for 3s to enter main menu Item Figure Function 1 Shows temperature unit 2 Shows ESD status ON OFF 3 Shows the unit status Lock Unlock 4 350 Display real time temperature 5 Show...

Страница 9: ...d setting temperature method Temperature setting 1 Press the button once to increase the value by one digit press and hold it to continuously increase 2 Press the button once to decrease the value by...

Страница 10: ...adjust the temperature value see picture 2 4 Press STORE to save it and press BACK to return Temp1 menu 6 3 Alarm Temperature Setting 1 Press ENTER button to enter into menu and blue data is shown 2...

Страница 11: ...in the ON status 4 Press STORE button to save it and press BACK button to return Alarm menu see picture Note The alarm temperature setting range is 2 99 6 5 Sleep time setting 1 Press ENTER button to...

Страница 12: ...The sleep time is reached the soldering station will power off automatically 6 6 Close time setting 1 Press ENTER button to enter into menu and blue data is shown 2 Press 1 button to select close mode...

Страница 13: ...it and press BACK to return unit menu see picture 6 8 Key Tone Setting 1 Press ENTER button to enter into Key Tone menu 2 Press 1 button to select sound mode ON or OFF 3 Press STORE button to save it...

Страница 14: ...shown 2 Press the or button to change the set value 3 Press the STORE button to save the setting value and press BACK button to return 6 11 Language selection 1 Press ENTER button to enter Language me...

Страница 15: ...r the default password 000000 Press and button to adjust password value and press 1 to move the cursor to the next digital 2 Press ENTER button to confirm it 3 Enter the new password twice 4 Press ENT...

Страница 16: ...the value is the same as the thermometer press STORE to calibrate it Tip Care and Maintenance Select Tip 1 Select a tip that maximizes contact area between the tip and solder joint Maximizing contact...

Страница 17: ...16 3 Tip type Soldering tip maintenance 1 High temperature would affect decrease the lifetime of soldering tip Set the...

Страница 18: ...tip with solder to prevent oxide 2 Set the temperature as per application but as low as possible 3 Choose the right type of soldering tip 4 The plating would be broken if the tip is bent Do not use th...

Страница 19: ...18 NO handle handle is not insert Sensor shorten sensor short circuit Heater broken heating element open Tip low temperature alarm...

Страница 20: ...QUICK INTELLIGENT EQUIPMENT CO LTD ADD NO 11 FengXiang Road Wujin High Tech Industrial Development Zone Jiangsu China TEL 86 519 86225678 FAX 86 519 86558599 POSTCODE 213167 WEBSITE www quick global c...

Отзывы: