9
7.2.1. Channel Settings
1) Press
key to select the channel bar, and continue to press
key to select the
parameter bar.
2) Press + or - key in the channel bar to select CH1; In the parameter bar, set the
temperature and press
key to save.
3) In CH3, select the temperature zone and press
key to enter the temperature
zone setting interface to realize the function of segmented temperature control. Two
temperature ranges can be set, in which the temperature and time can be adjusted.
The range of temperature 1 is 50
℃
- 200
℃
, the range of temperature 2 is 200 -
350
℃
, and the range of time 1 and 2 is 000 - 600 seconds.
When temperature 1 reaches the set temperature, the countdown starts (time 1 is set
according to requirements). When the countdown ends, it will enter temperature 2.
When temperature 2 reaches the set temperature, the countdown starts (time 2 is set
according to requirements). When the countdown ends, it will return to temperature 1
and remains there. If you want to continue the above functions, press
key in the
main interface to turn off and turn on again.
4)Press
key to go back level by level until you return to the main interface.