USER MANUAL
10
2.9.
Calibration
It is better to calibrate the temperature after changing the heater element.
Adopt two point temperature calibrate, the method of calibrating the
temperature is as following:
1.
In work state, set the calibration temperature 300
℃
and 500
℃
respective
calibrate.
When the temperature is stable and the window displays “Real Temp”, test
the outlet temperature of the handle with thermometer and write down it.
2.
Into the calibration mode: When the temperature is stable, press the keys
“CH1” and “CH3” keys at the same time about 15seconds, and then the LCD
will display “cal” about two seconds and the system will enter into the
temperature calibrating mode.
Input the testing temperature after the hundreds digit twinkling.
3.
Press “TEMP
▲
”or “TEMP
▼
” key to setting the data and press “AIR
▲
”or “AIR
▼
” key to move the cursor. After inputting the testing result,
click the “CH2” key.
4.
If calibration is successful, the LCD respectively displays “C-L/C-H”
,
if not
successful, the LCD displays “no”.
5.
If the temperature still has some departure, you can repeat calibration
according with the above steps.
NOTE: Suggest using 191 or 192 thermometer to measure the temperature.
2.10.
Heating Element
2.10.1. Steps of removing the heater