7.    CHANGING THE SENSOR

 

 

Note

While the case is open the unit is not intrinsically safe; perform this 
exercise in a non-hazardous area. Allow the new sensor to stabilize. 
Zero and calibrate prior to use.

 

 

The entire sensor module is replaced. The old sensor module should be 
discarded; do not attempt to break the module apart or reuse any parts.

 

 
To change the sensor, first remove the battery and its holder. Inside of the 
battery compartment is a single screw. Turning the screw counter-clock 
wise pushes up and releases the red top cap of the unit. Lift the sensor 
module up off of the circuit board and push in the new one. A circular 
hydrophobic filter sits in the opening on top of the sensor module; make 
sure the new sensor has a filter in place prior to reinstalling the top cap. Set 
the top cap back on the unit and turn the screw clockwise until the top cap 
again seals against the unit's main housing. When the battery is replaced the 
unit will turn on and read the new sensor information. 

 

 

 

8.    MAINTENANCE

 

 
When the unit is stored or not in use for extended lengths of time, it should be 
calibrated, functionally tested, and its battery should be checked or changed at 
least once per month. This will help to ensure that the unit will be working 
properly when it is needed.

 

 

Note

If the unit has been stored without a battery, then it requires a 
stabilization period, prior to calibration, once the battery has been 
replaced.

 

Содержание SafeLog 100

Страница 1: ...oni toring needs The SafeLog 100 is an extremely rugged lightweight single gas datalogging personal monitor It is our goal to make your decision to buy Quest products the right one and to provide supp...

Страница 2: ...Date 12 3 3 8 Temperature 12 3 3 9 Password 12 3 4 Zero Mode 12 3 5 Print Mode Reset 13 3 5 1 Print 13 3 5 2 Reset 13 3 6 Run Mode 13 3 7 Zeroing 13 3 8 Calibration 14 3 8 1 Minimum Flow Rate Table 16...

Страница 3: ...All recorded data can be sent to a printer or computer for record keeping or further data analysis Like many of Quest s data logging instruments the SafeLog 100 works with the QuestSuite software pac...

Страница 4: ...s level is exceeded the alarm activates and the alarm value is shown on the display 2 3 4 STEL This annunciator which stands for Short Term Exposure Limit is present only when using a toxic gas sensor...

Страница 5: ...t is also on during all alarming conditions 3 UNIT OPERATION 3 1 Turning the Unit ON Press ON OFF The following startup sequence will occur 1 All of the display segments appear simultaneously on the s...

Страница 6: ...he display For the TWA alarm toxics only HI and TWA will be on the dis play Pressing the enter key causes the first digit to flash Use the arrow and enter keys to adjust the level 3 3 5 Confidence Chi...

Страница 7: ...r adjusts for small baseline drifts which occur over time in electrochemical sensors Zero the unit by pressing the zero key from the display with four dashes or as one of the options available through...

Страница 8: ...and TWA alarm conditions are checked every minute If any alarm setting is exceeded the buzzer and LED turn on the value causing the alarm is displayed and the alarm condition and time of occurrence a...

Страница 9: ...o a computer or in parallel to a printer Serial data transmission requires the Quest cable 54 715 and par allel data transmission requires the Quest cable 56 875 With the cable plugged into the unit p...

Страница 10: ...4 3 Sample Printout...

Страница 11: ...ery voltage drops below the usable range dEAd appears in the display the buzzer activates for approximately 25 seconds then the unit turns itself off The 9 volt alkaline battery must be replaced prior...

Страница 12: ...e new one A circular hydrophobic filter sits in the opening on top of the sensor module make sure the new sensor has a filter in place prior to reinstalling the top cap Set the top cap back on the uni...

Страница 13: ...GAS SPECIFIC DATA Gas Range Reso lution Accur acy Drift Alarm High Alarm LOW Oxygen O2 0 30 0 1 5 1 23 5 19 5 Sensor accuracy listed by City Technology as a percentage of the reading Long term output...

Страница 14: ...o 140 F storage Humidity 0 to 95 R H non continuous non condensing Range 15 to 90 R H continuous non condensing Warranty Sensors 1 year electronics 1 year Cell Life Ammonia sensor 1 year in air or 2pp...

Страница 15: ...tion Kit 54 974 Two Cylinder Calibration Kit Includes two cylinder calibration case regulator specify which regulator and tubing 12 4 Regulators 54 971 0 5 1pm Regulator with 2 050 I D tubing 54 972 1...

Страница 16: ...ipping damage please direct your calls or letters to the attention of the Service Manager Quest Technologies 262 567 9157 or 800 245 0779 Office hours are from 7 AM to 6 PM Central Standard Time Monda...

Страница 17: ...1060 Corporate Center Drive Oconomowoc WI 53066 USA 800 245 0779 Fax 262 567 4047 http quest technologies com Printed in U S A 054 479 Rev B...

Отзывы: